Wuhan GDMK Electric Co., Ltd.

2nd Floor, Western Color Aluminum-Plastic Comprehensive Building, Huangjiahu Avenue, Transformer Oil Chromatography, Wes 430200

https://www.mktester.com/transformer-oil-tester/transformer-oil-chromatography.html

Product description:

This instrument absorbs the advanced technology of similar products at home and abroad, and can set up various parameters through the keyboard. This machine has a power down protection, over temperature protection, "0" C protection and gas protection, automatic electronic ignition, etc. it has the characteristics of stable and reliable performance, simple and reasonable structure, simple operation, high expansibility, etc.


Technical index:

Temperature control:

a. Chromatographic column chamber temperature:

a.1 Temperature control range: room temperature +8?~400?(Set the temperature increment 1?)

a.2 The precision of temperature control: It is±0.1?, which is below 200?;

It is±0.2?, which is above 200?.

a.3 Temperature gradient :It is less than 1% in the column effective region.

a.4 The deviation between indicating temperature and set temperature is less than 0.5 ?.

a.5 The deviation between actual temperature and indicating temperature is less than 2%

6 Maximum heating power 1500W

a.7 PT100 corundum ceramic platinum resistance is used in temperature sensing element

b. Convert temperature:

b.1 The precision of temperature control:

It is±0.1?, which is below 200?;

It is±0.2?, which is above 200?;

b.2 Maximum service temperature 400?

b.3 Adopt horizontal heating, two pieces of 100 W with internal heating types of stainless steel bar

b.4 PT100 corundum ceramic platinum resistance is used in temperature sensing element

c.Hydrogen flame detector temperature:

c.1 The precision of temperature control: It is±0.1?, which is below 200?;

It is±0.2?, which is above 200?;

Maximum service temperature 400?

c.2 Adopt horizontal heating, two pieces of 90 W with internal heating types of stainless steel bar

c.3 PT100 corundum ceramic platinum resistance is used in temperature sensing element

d.Thermal Conductivity detector temperature:

d.1The precision of temperature control: It is±0.1?, which is below 200?;

It is±0.2?, which is above 200?;

Maximum service temperature 400?

d.2 Adopt vertical circular heating, two pieces of 90 W with internal heating types of stainless steel bar

d.3 PT100 corundum ceramic platinum resistance is used in temperature sensing element

Thermal Conductivity detector

a. sensitivity: S=3000mv·ml/mg(benzene, H2)

b. noise: =0.02mv

c. drift: =0.1mv/h

d. built-in preamplification

e. Semi diffusion type, 100O four arm rhenium tungsten wire

f. Constant-current source power supply mode

(3) Hydrogen flame detector

a. Limit of detection M=2×10-11g/s (benzene/ carbon disulfide)

b. noise: =5×10-13A

c. drift: =5×10-12A/30min

d. Lift limit of detection: M=1×10-9g/s

e. Lift drift: every cycle=5×10-12A

f. whole collector type, metal nozzle

Platinum ignitor fuse

(4)Dimension and weight

a. Main machine dimension: 610(width)*460(height)*470(depth)

b. Weight: about 60kg

3. Instrument optional peripheral equipment and accessories:

(1)Workstation of chromatography data.

(2)Gas source:

hydrogen, air,nitrogen(99.999%).

(3)planar six-way gas sampler

(4) Methane conversion oven